RPS-160-12 Series

RPS-160-12 Series Features
Universal AC input
Built-in active PFC function
Short circuit, over load, over voltage, over temperature protections
5″x3″ compact size
Free air convection for 110W and 160W with 20.5 CFM forced air
With power good and fail signal output
Built-in remote sense function for 5~15V
No load power consumption under 0.75W by PS-ON control
Standby 5V@1A with Fan/0.6A without Fan
Approval: UL / CUL / TUV / CE / CB
Part Number Voltage Current (amps) Power (watts)
RPS-160-12 12V 13A 112.2W

Specifications
Input Voltage                                       90 ~ 264Vac 127 ~ 370Vdc
Input Frequency                                  47 ~ 63Hz
Power Factor                                       PF>0.93/230Vac PF>0.98/115Vac at full load
Input Current                                       2A/115Vac 1.1A/230Vac
Inrush Current                                     35A/115Vac 70A/230Vac, cold start
Leakage Current                                 Earth leakage current <300uA/ 264Vac, Patient leakage current                                   <100uA/264Vac
Line Regulation                                   ±0.5%
Load Regulation                                  ±1.0%
Set Up, Rise Time                               1200ms, 30ms/230Vac 2500ms, 30ms/115Vac at                                                                  full load
Hold Up Time                                      16ms/230Vac/115Vac at full load
Over Load Protection                          105 ~ 135% rated output power