| Applications Non-patient contact medical General electronics IT equipment Industrial  | 
|
![]()  | 
Features Universal AC input/full range Ultra low profile Isolation class II No load power consumption <0.5W Fixed switching frequency at 67KHz Short circuit, OLP, OVP & OTP Compliance to UL60601-1 & TUV EN60601-1 medical standards Cooling by free air convection  | 
| Part Number | Voltage | Current (amps) | Power (watts) | 
| NFM-10-3-3 | 3.3V | 2.5A | 8.25W | 
| NFM-10-5 | 5V | 2A | 10W | 
| NFM-10-12 | 12V | 1A | 10.2W | 
| NFM-10-15 | 15V | 1A | 10.05W | 
| NFM-10-24 | 24V | 0.42A | 10.08W | 
Specification
Input Voltage                                    85 ~ 264VAC 120 ~ 370VDC
Input Frequency                               47 ~ 440Hz
AC Current                                       0.25A/115VAC 0.15A/230VAC
Inrush Current                                  25A/115VAC 45A/230VAC typically, cold start        
Setup, Rise Time                             1000ms, 20ms/230VAC 1000ms, 20ms/115VAC at full load
Hold up Time                                   100ms/230VAC 20ms/115VAC typically at full load
Overload Protection                         Above 105% rated output power
Over Voltage Protection                   Protection type shut off
Over Temperature Protection          Tj 145°C typically (U1) Detect on main control IC
                                                         Shut down o/p voltage, recovers automatically after temperature goes down
Working Temperature                      -20°C to +70°C (Refer to output load de-rating curve)
Working Humidity                            20 ~ 90% RH non-condensing
Withstand Voltage                           I/P-O/P:4KVAC
Isolation Resistance                        I/P-O/P:100M Ohms/500VDC
Description
Featuring class II isolation levels they can be equipped into non-patient contact medical equipment and still comply with class B as per EMI regulations. With energy saving and the environment in mind the no load power consumption is lower than 0.5W so that it complies with the requirements for “green mode” power supplies. Standard functions include short circuit; over load and over voltage protection and over temperature protection is equipped on the whole NFM series. Suitable for use in general medical or for industrial usage that require low leakage current, general electronic applications and low profile IT equipment this on-board type PSU’s can be applied on the main control PCB of the final systems.